Instrument

29-1.jpg29-2.jpg

 

 

MYA 3Y microbalances are designed to meet the highest requirements on accuracy of mass measurements. Series F is dedicated to weighing filters. Measurement reliability and accuracy are maintained by automatic system of internal adjustment / calibration.
Microbalances comprise two major components (an electronic module and a precise mechanical measuring system are enclosed separately). Such design eliminates the influence of heat sourcing from instrument's electronics on its mechanical components and additionally protects it from shocks and vibrations caused by users operating the instrument.
All the elements of a microbalance are made of glass and steel which eliminates the influence of electrostatics on weighing process.
기능(Functions)
  - Filling
  - Checkweighing
  - Percent setup
  - Statistics
  - Formulation
  - GLP procedures
  - Autotest
  - Infrared sensors
  - 3 years warranty period
특징(Features)
  Microbalance Weighing Chamber
  Microbalance Weighing Chamber is equipped with openable pan intended for filter weighing
   
  Communication interfaces
  - Ethernet
  - 2×RS 232
  - 2×USB
  - additional display port
   
  Infrared proximity sensors
  Optional functions:
  - PRINT function
  - TARE function
  - sensor's sensitivity adjustment
   
  Electronic level indicator
  Functions:
  - ALARM function
  - graphic level indicator
  - programmable acceptable tilts
   
  Data exchange through USB storage devices
  - update balance software
  - export weighing data
  - export/import databases
  - export/import balance settings
  - exchang data between balances
   
  Additional weighing pan ø 60 mm
  - designed for weighing filters
  - used interchangeably on a microbalance MYA 0,8/3
Technical data: MYA 0,8/3.3Y MYA 5.3Y.F MYA 5.3Y.F1
Max capacity 0,8/3 g 5 g
Readability 1/10 µg 1 µg
Tare range -3 g -5 g
Repeatability * 1 µg 1,6 µg
Linearity ±3 µg ±5 µg
Eccentric load deviation 3 µg 5 µg
Sensitivity offset 1,5×10-6×Rt
Sensitivity temperature drift 1×10-6/°C×Rt
Sensitivity stability 1×10-6/Year×Rt
Minimum weight (USP) 3 mg 4,8 mg
Minimum weight (U = 1%, k = 2) 0,2 mg 0,32 mg
Pan size ø 16 + ø 60 mm
(weighing pan for filters)
ø100 mm + ø 26 mm ø160 mm + ø 26 mm
Weighing chamber dimensions ø 90×90 mm ø118×35 mm ø168×35 mm
Stabilization time 5 s
Adjustment/calibration internal (automatic)
Power supply 13,5 ÷ 16 V DC / 2,1 A
Display 5,7'' touch screen
Interface 2×USB, 2×RS 232, Ethernet, 2Inputs/2Outputs
Net weight/Gross weight 10,2/14,7 kg
Packaging size 565×565×355 mm
Ambient conditions:
Working temperature +10° ÷ +40°C
Atmospheric humidity ** 40% ÷ 80%
PLUS