Instrument

27.jpg

 

 

특징(Features)
  - 99% 이상의 유기물 , 무기물, 미립자 제거
  - LCD monitor를 이용한 간단한 사용법
  - 지속적인 Quality monitoring : 사용자가 설정한 수질 이하의 일차수 생산시 경보 작동
  - RS232포트를 이용한 관련 데이타의 전송 및 저장 (GLP 응용에 적합)
  - 살균 포트가 있어서 손쉽게 시스템을 살균할 수 있음
  - Type III primary-grade water from a potable tap feed
  - Water pre-treatment prior to ultrapure water purification
  - Routine glassware washing
  - Hydroponics (수경재배)
  - Feed water for humidifiers and autoclaves (Autoclave와 가습기의 공급수)
  - Use in environmental cabinets and stills
  - Steam generators
Potable water enters through a strainer or is passed through the pre-treatment cartridge. The pre-treatment cartridge has been designed to protect the reverse osmosis cartridges from particulate/colloidal matter and excessive free chlorine that may be present in the incoming feed water.
The water passes the sanitization port and moves through two reverse osmosis cartridges, set up in series, which split the flow into permeate and concentrate streams. The permeate water is further purified while the waste concentrate stream is passed to drain.
The permeate water is passed through the:
1. Water quality sensor, which measures the conductivity of the water
2. Temperature sensor, which provides accurate temperature measurement

To ensure water quality is maintained at the highest levels, the unit has a built-in, automated auto-rinse cycle. This cycle is performed each time the process is initiated and consists of a 1-minute high flow rinse across the RO cartridges to the drain.
- Cascada BIO-water Purification System: Faster, consistent water purification with real-time quality monitoring for life sciences applications
- Cascada AN-water Purification System: Faster, consistent water purification with real-time quality monitoring for analytical applications
- Cascada LS-water Purification System: For life sciences applications at an economical price
- Cascada IX-water Purification System: For general laboratory applications at an economical price
- Cascada Purification Systems - Accessories and Replacement Parts
Water Purity RO-water
Inorganics - min. > 90% rejection
Inorganics - typical Up to 98% rejection
Organics (MW > 100 Dalton) > 99% rejection
Total organic carbon (TOC) - typical
(dependent on feed water)
< 0.1 ppm
Bacteria < 5 CFU/mL
Particles > 99% rejection
Flow Rates 15 L/hour
Safety Features RO-water
Power fail-safe Yes
Automatic level controls Yes
Audio-visual alarm warnings Yes
Auto-restart Yes
System Monitoring RO-water
RO permeate purity Yes
RO permeate temperature Yes
Reservoir level Yes
Consumable remaining life indicator Yes
규격(Specifications)
  Dimensions
  (without docking vessel)
41 W x 52 H x 26 D cm (16.2 W x 20.5 H x 10.5 D in.)
  (with docking vessel)
57 W x 74 H x 32 D cm (22.4 W x 29.1 H x 12.6 D in.)
   
  Weight
  (without docking vessel)
13.0 Kg (28.5 lbs.)
  (with docking vessel)
16.5 Kg (36.2 lbs.)
  Electrical Requirements
Electrical Input: 100 - 240 V AC, 50 - 60 Hz
System Voltage: 24V DC
  Feedwater Pressure
Maximum: 2.0 bar (200 kPa, 30 psi)
Minimum: Flooded suction
  Operating Temperature
1 - 35 °C (34 - 95 °F)
Conforms to ASTM Type III, NCCLS Type III, CAP, and BS3978 Grade 3 specifications.
   
  Water Qualty Sensor - 물의 전도율 측정
  Temperature Sensor- 정확한 온도 측정
  Pre0treatment (PAL-LC140) - 전처리 필터로 RO cartridge 보호
Part Number Description Pkg
PAL-CAXXXROM2 Cascada RO-water system 1/pkg
PAL-C176 PURIFICATION (CID) CARTRIDGE PACK 1/pkg
PAL-C173 UV Lamp 185um 1/pkg
PLUS